User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Meriam Meri-Cal DP200I:Digital Pressure Meter DATE: 17-Dec-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 13 NUMBER OF LINES: 41 CONFIGURATION: Fluke 5700A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P F W 1.002 ASK+ X 1.003 HEAD INITIAL CONDITIONS 1.004 DISP Connect the 5700A OUTPUT HI/LO to the 1.004 DISP UUT VOLTS/COM terminals. 1.005 DISP Press any key on the UUT to turn it on. 1.006 DISP Let UUT warm up for 15 minutes. 1.007 5700 0V S 2W 1.008 DISP Wait for the UUT reading stabilize, then press 1.008 DISP the ZERO SET button followed by the ENT button. 1.009 HEAD {* Volts Calibration *} 1.010 JMP 2.001 1.011 EVAL 2.001 5700 30.000V 0.1% 0.02U 2W 3.001 5700 20.000V 0.1% 0.02U 2W 4.001 5700 10.000V 0.1% 0.02U 2W 5.001 5700 3.0000V 0.1% 0.002U 2W 6.001 5700 1.0000V 0.1% 0.002U 2W 7.001 HEAD {* Current Calibration *} 7.002 JMP 8.001 7.003 EVAL 8.001 DISP Move the connection to the UUT VOLTS jack 8.001 DISP to the UUT mAMPS jack. 8.002 5700 30.000mA 0.1% 0.02U 2W 9.001 5700 20.000mA 0.1% 0.02U 2W 10.001 5700 10.000mA 0.1% 0.02U 2W 11.001 5700 3.0000mA 0.1% 0.002U 2W 12.001 5700 1.0000mA 0.1% 0.002U 2W 13.001 END